Référence Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 436, 3, page (313-325)
Publication Publié, 1999
;Bouhali, Othmane
;Stefanescu, Jacqueline
;Vander Velde, Catherine
;Vanlaer, Pascal
;Van Lancker, Laurent H.C.-P.
; [et al.]Référence Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 436, 3, page (313-325)
Publication Publié, 1999
Article révisé par les pairs
| Résumé : | In order to check the system aspects of the forward-backward MSGC tracker designed for the future CMS experiment at LHC, 38 trapezoidal MSGC counters assembled in six multi-substrates detector modules were built and exposed to a muon beam at the CERN SPS. Results on the gain uniformity along the wedge-shaped strip pattern and across the detector modules are shown together with measurements of the detection efficiency and the spatial resolution. |