par Aiello, Sebastiano;Colomer Molla, Marta
Référence Journal of Instrumentation, 13, 5, P05035
Publication Publié, 2018-05-01
Article révisé par les pairs
Résumé : The Hamamatsu R12199-02 3-inch photomultiplier tube is the photodetector chosen for the first phase of the KM3NeT neutrino telescope. About 7000 photomultipliers have been characterised for dark count rate, timing spread and spurious pulses. The quantum efficiency, the gain and the peak-to-valley ratio have also been measured for a sub-sample in order to determine parameter values needed as input to numerical simulations of the detector.