par Huberland, Francois ;Rio-Tinto, Ricardo;Cauche, Nicolas ;Dugardeyn, Sonia ;Delattre, Cecilia ;Sandersen, Charlotte;Rocq, Laureen ;Van Ouytsel, Pauline;Delchambre, Alain ;Devière, Jacques ;Blero, Daniel
Référence Endoscopy
Publication Publié, 2021-07-01
Référence Endoscopy
Publication Publié, 2021-07-01
Article révisé par les pairs
Résumé : | Background A medical device that allows simple and safe performance of an endoscopic septotomy could have several applications in the gastrointestinal (GI) tract. We have developed such a device by combining two magnets and a self-retractable wire to perform a progressive septotomy by compression of the tissues. We describe here the concept, preclinical studies, and first clinical use of the device for the treatment of symptomatic epiphrenic esophageal diverticulum (EED). Methods The MAGUS (MAgnetic Gastrointestinal Universal Septotome) device was designed based on previous knowledge of compression anastomosis and currently unmet needs. After initial design, the feasibility of the technique was tested on artificial septa in pigs. A clinical trial was then initiated to assess the feasibility and safety of the technique. Results Animal studies showed that the MAGUS can perform a complete septotomy at various levels of the GI tract. In two patients with a symptomatic EED, uneventful complete septotomy was observed within 28 and 39 days after the endoscopic procedure. Conclusions This new system provides a way of performing endoluminal septotomy in a single procedure. It appears to be effective and safe for managing symptomatic EED. Further clinical applications where this type of remodeling of the GI tract could be beneficial are under investigation. |