par Scholer, André;Boecker, Jörg;Engelmayer, Ulf;Feldmann, Knut;Hannak, Dieter;Kattermann, Reinhard;Oellerich, Michael;Raith, Hannelore;Schlebusch, Harald;Wieland, Heinrich;Willems, Dominique
;Jarausch, Jochen;Domke, Ingrid
Référence Clinical chemistry, 43, 1, page (92-99)
Publication Publié, 1997
;Jarausch, Jochen;Domke, IngridRéférence Clinical chemistry, 43, 1, page (92-99)
Publication Publié, 1997
Article révisé par les pairs
| Résumé : | A new turbidimetric inhibition immunoassay for digoxin (Tina-quant® [a] Digoxin, Boehringer Mannheim) was evaluated in seven laboratories. It can be performed without sample pretreatment with ready-to-use reagents on nondedicated analyzers in combination with routine clinical chemistry. The studies revealed a good analytical performance: lower limit of detection 0.12 μg/L (3 SD from mean of blank); linearity up to 7.5 μg/L; median between- run CVs 8.1% (0.6 μg/L), 2.8% (1.5 μg/L), 1.9% (3 μg/L); mean analytical recovery in control sera 98-102%; slopes from 0.97 to 1.09 and intercepts from -0.28 to 0.10 μg/L in comparison with four immunoassays; and a high resistance to common interferents. The test was more resistant to digoxin- like immunoreactive factor (DLIF) interference than other methods, showing cross-reactivity only in some intensive care patient samples. Among 192 patients in whom DLIF is expected (e.g., pregnant women, patients with renal failure, newborns), 90% of results were ≤0.26 μg/L digoxin. Cortisol showed no cross-reactivity and digoxigenin had a low reactivity. An interlaboratory survey revealed a good comparability of the Tina-quant [a] test with the median of all methods (slope 0.99, intercept -0.06 μg/L). An HPLC method for digoxin based on isocratic separation of samples on an RP-18 column followed by detection by an immunoassay yielded a reasonable comparability with the immunochemical tests with noncritical samples. Divergent results of immunoassays caused by DLIFs or different cross-reactivities with digoxin metabolites or derivatives can be explained by the use of this HPLC method. |