par De Tiege, Xavier ;de Beeck, Marc Op;Funke, Michael;Legros, Benjamin ;Parkkonen, Lauri;Goldman, Serge ;Van Bogaert, Patrick
Référence Epilepsy research, 82, 2-3, page (227-231)
Publication Publié, 2008-12
Référence Epilepsy research, 82, 2-3, page (227-231)
Publication Publié, 2008-12
Article révisé par les pairs
Résumé : | Ten patients with focal epilepsy were studied with magnetoencephalography (MEG) to determine if a new light-weight magnetically shielded room (lMSR) provides sufficient attenuation of magnetic interference to detect and localize the magnetic correlates of epileptic activity. Interictal MEG epileptic events co-localizing with the presumed location of the epileptogenic zone were found in all patients. MEG measurements performed in the lMSR provide an adequate signal-to-noise ratio for non-invasive localization of epileptic foci. |